Background
The Z1 CPAP device(model: HD60-1007, HD60-1017) is unlike its predecessors in that it’s extremely lightweight and can easily be used for travel. It includes a multitude of features such as Z-Breathe exhalation, automatic altitude adjustment, and leak compensation. The Z1 is designed to be the most portable machine available. Because of that, there is no dedicated humidifier option offered for the Z1.
Default Settings
- Ramp: 15 Min
- Ramp Start Pressure: 4 cm H2O
- ZBreathe Setting: CPAP 2
- APAP Ranges: 4 - 20 cm H2O
Key Features
- Small and Lightweight Automatic Travel Machine
- Built-in Bluetooth for Nitelog iPhone App
- Sophisticated Data Management
- Z-Breathe Pressure Relief
- Leak Compensation and Altitude Adjustment
- Innovative Ramp Feature
- Quiet Operation
- Optional Battery Available
Identification
The Z1 CPAP device is identifiable by the name on the front. However, if you are looking for a specific model number, you can find it simply on the back of the machine. There is a screen with three buttons on the front and two ports. The big port in the center is where the CPAP tube should go, and the other is for the charger.
Technical Specifications
Operating pressure range 4 to 20 cm H2O Maximum single fault steady pressure state 30 cm H2O Deliverable pressure accuracy ±0.6 cm H2O ± 4% of the measured reading
Unit Input range 100–240V, 50–60Hz Typical power consumption 20W (30VA) Maximum power consumption 35W (70VA) Operating temperature 5ºC to 35ºC (41ºF to 95ºF) Operating humidity 10–80% Relative Humidity, Non-Condensing Operating altitude Sea level to 2400 m (8000 ft) Operating atmospheric pressure 101 to 75.3 kPa Storage and transport temperature -20ºC to +60ºC (-4ºF to +140ºF) Storage and transport humidity 10-95% Relative Humidity, Non-Condensing Storage and transport altitude No limit Storage and transport atm. pressure No limit Electromagnetic compatibility Compatible with IEC 60601-1-2 Aircraft use Complies with RTCA DO-160G Standard for use in flight IEC 60601-1 classification Class II (double insulation), Type BF Available modes Standby, CPAP, Auto-CPAP and CPAP/Auto-CPAP with Ramp Deliverable pressure accuracies ±0.6 cm H2 O ± 4% of the measured reading Sound pressure level < 26 dBA Nominal dimensions (L x W x H) 16.46 cm x 8.38 cm x 5.13 cm (6.48” x 3.30” x 2.02”) Weight 284 g (10 oz.) Housing construction Flame retardant engineered thermoplastic Supplemental oxygen Not for use with oxygen Standard air filter Polyester Air outlet The 22 mm air outlet complies with ISO 5356-1 PowerShell Battery Specifications 99.4 Watt-hours capacity Lithium Ion Battery Pack, 4S2P Weight approx. 440 g (15.5 oz.)
Expected Service Life Z1, PowerShell and Power Supply: 5 years Tube Adapter, USB Cable, and End Cap: 2 years
Unit Input range 100–240V, 50–60Hz Typical power consumption 20W Maximum power consumption 35W Operating temperature 5ºC to 35ºC Weight 284 g Lithium Ion Battery Pack
Z breath setting: 3
Treatment Pressure: 20cm H2O
Operating Time: 8.9 Hours